You will be leaving the Cook Medical website that you were viewing and going to a Cook Medical website for another region or country. Not all products are approved in all regulatory jurisdictions. The product information on these websites is intended only for licensed physicians and healthcare professionals.

Blue Rhino® G2-Multi Percutaneous Tracheostomy Introducer

Welcome History Benefits of PDT Products Tracheostomy Tubes Educational Resources

Educational resources


Resources

Prepping for the procedure

6307577676112
brightcove
true

Instructions for Use

Procedural guide

For complete prescribing information, including indications for use, warnings, precautions, adverse events, and procedural instructions, refer to the Instructions for Use (IFU), which can be found on the Blue Rhino G2-Multi product page.

Patient preparation

Procedural guide(Fig. 1)
1. Following the tracheostomy tube manufacturer’s instructions, test the balloon cuff and inflation system.
2. Place the patient in the tracheostomy position. (Fig. 4) Position a pillow under the shoulders to permit full extension of head and neck. The head of the patient‘s bed may be elevated 30-40 degrees at the physician‘s discretion.
3. Use ventilator changes and sedation to control patient respiration. A PEEP (positive end expiratory pressure) level of 5-10 is recommended.
Procedural guide(Fig. 4)
4. Instruct the clinician managing the airway to loosen the fixation tapes of the in-place endotracheal tube and deflate the cuff, making necessary changes in tidal volume, frequency, etc., to evaluate compensation needed for air leak. Reinflate the endotracheal tube cuff. Continuous oximetry monitoring should be employed.
Procedural guide(Fig. 2)
5. Prep and drape the anterior neck area.

6. Generously lubricate the surface of the appropriately sized loading dilator and load the tracheostomy tube onto the dilator. Ensure that the tracheostomy tube’s tip fits snugly on the dilator. (Fig. 2) Ensure that the balloon is completely deflated. Thoroughly lubricate the tracheostomy tube assembly.

Tracheostomy preparation

Procedural guide(Fig. 3)
NOTE: Dual cannula tracheostomy tubes may also be placed using the following technique. The inner cannula must be removed for introduction. Always check the fit of the dilator to the tracheostomy tube prior to insertion.

1. Palpate landmark structures (thyroid notch, cricoid cartilage) to ascertain the proper location for tracheostomy tube placement. Access and, ultimately, tube placement is ideally made at the level between the first and second tracheal cartilages or between the second and third tracheal cartilages whenever feasible. (Fig. 3)

Procedural guide(Fig. 5)
2. After introducing local anaesthesia, make a 1.5-2.0 cm skin incision (vertical or horizontal) at the chosen insertion site. (Fig. 5)
Procedural guide(Fig. 6)
3. If desired, use a curved mosquito clamp to gently dissect vertically and transversely down to the anterior tracheal wall. (Fig. 6) With a finger tip, dissect the front of the trachea, in the midline, free of any tissues and identify the cricoid cartilage. Displace the isthmus of the thyroid downward, if present. NOTE: An adequate skin incision and blunt dissection of the subcutaneous tissue can minimise the need for excessive force and torque throughout the procedure. Excessive force and rotation may lead to long-term complications (e.g., stenosis).

4. Deflate the endotracheal tube cuff and withdraw to an appropriate distance above the insertion site, yet still within the trachea. Re-inflate the cuff once the proper position of the endotracheal tube has been reached.

Procedural guide(Fig. 7)
5. Attach a syringe half-filled with fluid to the introducer needle and seek the tracheal air column by directing the needle, in the midline, posterior. Verify the needle’s entrance into the tracheal lumen via aspiration on the syringe resulting in air bubble return. (Fig. 7) Alternatively, if using bronchoscopy, visualise the needle entering the trachea. NOTE: It is important that the needle not impale the endotracheal tube. To ensure that the endotracheal tube is not impaled, gently move it in and out 1 cm. If the tube is impaled, the needle will be seen and felt to also move. If this occurs, it will be necessary to withdraw the needle, pull back the endotracheal tube, and then reinsert the needle. NOTE: Proper positioning and alignment may help minimise complications (e.g., stenosis).

6. With the needle tip positioned in the trachea, local anaesthesia may be injected (if necessary).

7. When free flow of air is obtained, with no impalement of the endotracheal tube, remove the inner needle of the introducer needle assembly and advance the outer FEP sheath several millimetres. NOTE: If using an introducer needle without a sheath, proceed to step 9.

Procedural guide(Fig. 8)
8. Attach a syringe to the FEP sheath and re-confirm position within the tracheal lumen by visualising free flow of air into the syringe when aspirated. (Fig. 8) Alternatively, re-confirm position by visualising the FEP sheath in the trachea with the bronchoscope. Remove the syringe.
Procedural guide(Fig. 9)
9. Introduce the J-tipped wire guide several centimetres into the trachea. (Fig. 9) NOTE: The wire should advance freely, without resistance. If resistance is encountered, do not force the wire guide. Confirm correct FEP sheath or introducer needle placement via bronchoscopy, then advance the wire guide into the trachea until the distal mark on the wire guide reaches skin level.

10. Remove the FEP sheath or introducer needle while maintaining wire guide position within the tracheal lumen. (Fig. 9)

Procedural guide(Fig. 10)
11. Maintaining the wire guide‘s position at the skin level mark, dilate the initial tracheal access site by advancing the short, 14 French introducer dilator over the wire guide with a slight twisting motion. (Fig. 10)

12. Remove the dilator while maintaining wire guide position.

13. Activate the hydrophilic coating by immersing the distal end of the Blue Rhino G2-Multi dilator in sterile water or saline.

Procedural guide(Fig. 11)
14. Advance the Blue Rhino G2-Multi dilator and the guiding catheter as a unit over the wire guide, while maintaining wire guide position. (Fig. 11) NOTE: Align the proximal end of the guiding catheter at the mark on the proximal portion of the wire guide. (Fig. 1) This will ensure that the distal end of the guiding catheter is properly positioned back on the wire guide, preventing possible trauma to the posterior tracheal wall during subsequent manipulations. NOTE: Bronchoscopic guidance may also prevent possible trauma to the posterior tracheal wall.

15. Begin to dilate the tracheal access site by advancing the guiding catheter and Blue Rhino G2-Multi dilator into the trachea. To properly align the dilator on the wire guide/guiding catheter assembly, position the proximal end of the dilator at the single positioning mark on the guiding catheter. This will ensure that the distal tip of the dilator is properly positioned at the safety ridge on the guiding catheter to prevent possible trauma to the posterior tracheal wall during introduction. While maintaining the visual reference points and positioning relationships of the wire guide, guiding catheter and dilator, advance them as a unit to the skin level mark on the Blue Rhino G2-Multi dilator. (Fig. 11) NOTE: Proper positioning and alignment may help minimise complications (e.g., stenosis).

Procedural guide(Fig. 12)
16. Advance and pull back the dilating assembly several times to effectively dilate the tracheal access site. NOTE: The wire guide must always lead the dilator and the guiding catheter assembly to prevent possible trauma to the posterior tracheal wall during dilation. Care should be taken to keep the guiding catheter assembly properly aligned with the mark on the proximal portion of the wire guide. This will ensure that the tip of the guiding catheter assembly does not advance beyond the distal tip of the wire guide within the trachea.

17. Remove the Blue Rhino G2-Multi dilator, leaving the wire guide/guiding catheter assembly in position. Respiratory air leak through the tracheostomy stoma should be noted to confirm intratracheal location of the wire guide and guiding catheter.

18. Advance the tracheostomy tube (loaded on the dilator) over the wire guide/guiding catheter assembly to the safety ridge of the guiding catheter, then advance wire guide, guiding catheter, loading dilator, and tracheostomy tube as a unit into trachea. (Fig. 12) NOTE: The assembly should be directed perpendicular to the axis of the trachea during insertion for uniform dilation between tracheal cartilages. Once the tracheostomy tube is within the tracheal lumen, the assembly may be directed caudad. NOTE: Proper positioning and alignment may help minimize complications (e.g., stenosis).

Procedural guide(Fig. 13)
19. Advance the tracheostomy tube to its flange.
Remove the dilator, guiding catheter, and wire guide, leaving the tracheostomy tube in place. (Fig.13) NOTE: At this point, the bronchoscope may be inserted into the tracheostomy tube to confirm correct placement. NOTE: If using a dual cannula tracheostomy tube, insert the inner cannula at this point.

20. Inflate the tracheostomy tube balloon cuff. Connect the tracheostomy tube to the ventilator. Confirm position of the tracheostomy tube via standard methods (e.g., capnography, breath sounds, etc.).

21. Deflate and remove the endotracheal tube.

22. Perform suction to determine if any significant bleeding or possible obstruction exists that has not been noted to this point.

23. If necessary, one suture may be taken at the bottom of the initial incision.

The effectiveness of percutaneous tracheostomy

Published articles

Bowen CP, Whitney LR, Truwit JD, et al. Comparison of safety and cost of percutaneous versus surgical tracheostomy. Am Surg. 2001;67(1):54–60.

Cabrini L, Landoni G, Greco M, et al. Single dilator vs. guide wire dilating forceps tracheostomy: a meta-analysis of randomised trials. Acta Anaesthesiol Scand. 2014;58(2):135–142.

Cabrini L, Pintaudi M, Winterton D, et al. Choice of the appropriate tracheostomy technique. In: Servillo G, Pelosi P, eds. Percutaneous Tracheostomy in Critically Ill Patients. New York, NY: Springer; 2016:67–78.

Cobean R, Beals M, Moss C, et al. Percutaneous dilatational tracheostomy: a safe, cost-effective bedside procedure. Arch Surg. 1996;131(3):265–271.

Delaney A, Bagshaw S, Nalos M. Percutaneous dilatational tracheostomy versus surgical tracheostomy in critically ill patients: a systematic review and meta-analysis. Crit Care. 2006;10(2):R55.

Freeman BD, Isabella K, Cobb JP, et al. A prospective, randomized study comparing percutaneous with surgical tracheostomy in critically ill patients. Crit Care Med. 2001;29(5):926–930.

Kluge S, Baumann HJ, Maier C, et al. Tracheostomy in the intensive care unit: a nationwide survey. Anesth Analg. 2008;107(5):1639–1643.

Kornblith LZ, Burlew CC, Moore EE, et al. One thousand bedside percutaneous tracheostomies in the surgical intensive care unit: time to change the gold standard. J Am Coll Surg. 2011;212(2):163–170.

Marra A, Danzi M, Vargas D, et al. Tracheostomy in intensive care unit: the need of European guidelines. In: Servillo G, Pelosi P, eds. Percutaneous Tracheostomy in Critically Ill Patients. New York, NY: Springer; 2016:155–159.

Mehta C, Mehta Y. Percutaneous tracheostomy. Ann Card Anaesth. 2017;20(Suppl 1):S19–S25.

Newhouse E, Ondik MP, Carr M, et al. Who is performing percutaneous tracheotomies? Practice patterns of surgeons in the USA. Eur Arch Otorhinolaryngol. 2011;268(3):415–418.

Rashid AO, Islam S. Percutaneous tracheostomy: a comprehensive review. J Thorac Dis. 2017;9 (Suppl 10):S1128–S1138.

Vargas M, Servillo G, Arditi E, et al. Tracheostomy in intensive care unit: a national survey in Italy. Minerva Anestesiol. 2013;79(2):156–164.

Thank You for Your Interest in Cook Medical's Products

The product information on this website is intended only for physicians and healthcare professionals licensed in the European Union (except France), the United Kingdom, Switzerland, Norway, Iceland, Turkey, and Liechtenstein. If you are located in another global region, please click on the regional flag at the very top of the webpage and choose your region from the drop-down options.